TGA Action ID RC-2025-RN-00579-1
Type of Product Medical Device
Product Name/Description Visions PV.018 Digital IVUS Catheter

Product number: 86700, 86700J
ARTG Number 337414
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has been informed of cases where physicians, during “radial-to-peripheral” procedures (radial access), have used a Digital IVUS Catheter without the appropriate sheath and/or guide catheter; this inappropriate use could lead to failure to provide adequate support for manoeuvring the guidewire and catheter to peripheral vessels.

When using a short radial artery sheath with a Visions PV.014P Digital IVUS Catheter over an .014 guide wire, for example, the user may experience resistance while loading and removing the catheter. If the user removes the wire and catheter together without forcing, there should be no expected harm to the patient.

However, if the user continues the procedure despite resistance, the catheter and guide wire could get tangled, potentially requiring further intervention. In some cases, longer sheaths can assist, but surgical removal may be necessary.

There have been 7 adverse events requiring additional intervention. No deaths have been reported for this problem.
Instructions for Customers/Users Customers/users can continue the use of Visions PV.018 Digital IVUS Catheters in alignment with the Instructions for Use. As per standard practice, take precautions when advancing or removing a catheter in complex vessel anatomies. Do not force a catheter into a narrow vessel or tight stenosis.

Philips will modify the Instructions for Use for Visions PV.018 Digital IVUS Catheters to include the new precaution: “Use a guide sheath of appropriate length to provide adequate support to the rapid exchange IVUS catheter and guidewire.”

The updated IFU will be available by the end of December 2025.

This action has been closed out on 06/10/2026
Action Commencement Date 2/07/2025
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team