TGA Action ID RC-2025-RN-00566-1
Type of Product Medical Device
Product Name/Description ROTEM sigma complete and ROTEM sigma complete + hep. An in vitro diagnostic medical device (IVD)

ROTEM sigma complete
Part Number: 555501
Lot: S240507, S240603, S240701, S240707, S240709

ROTEM sigma complete + hep
Part Number: 555502
Lot: S240607, S240608, S240706, S240708, S240804
ARTG Number 323925
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Werfen Global has identified that several lots of ROTEM sigma cartridges have a higher-than-expected likelihood of generating pressure errors during the ROTEM sigma functional checks.

These functional checks take place before and after the screen notification to add a sample. If this pressure error is identified after the sample has already been inserted, it will require sample recollection.

This action was undertaken prior to TGA notification.
Instructions for Customers/Users Customers were asked to identify affected lots and return them to be replaced at no charge.

This action has been finalised on 18/7/2025.
Action Commencement Date 24/06/2025
Responsible Entity Werfen Australia Pty Ltd