TGA Action ID RC-2025-RN-00565-1
Type of Product Medical Device
Product Name/Description Guedel Airway

One-piece Guedel airway, size 2, ISO 8.0, green
One-piece Guedel airway, size 3, ISO 9.0, yellow

REF / Product Code: 1112080
Lots: 32415284, 32416250, 32420438

REF / Product Code: 1113090
Lots: 32421079, 32422296
ARTG Number 224158
Unique Device Identifier (UDI) 5030267050659, 5030267050680
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action A non-conformance was raised for plastic burrs found inside the Guedel Airway during manufacture. The fault was caused by damaged production equipment contacting the inside of the Guedel Airway.
If the burr becomes detached and is inhaled, it could result in potential complications such as airway obstruction, tissue irritation, inflammation and infection.

The affected lot numbers have been distributed to hospitals and health services since the 4th of February 2025.

No reports of this problem have come from the market to-date.
Instructions for Customers/Users Customers are to immediately quarantine affected stock of listed lot numbers on hand to prevent further use and return it to Intersurgical Australia.

Replacement stock of unaffected lot numbers is available. Customers should contact customerservices@intersurgical.com.au to arrange replacement.

This action has been closed out on 08/04/2026