TGA Action ID RC-2025-RN-00564-1
Type of Product Medical Device
Product Name/Description IDS Ostase BAP EIA 96w. An in vitro diagnostic medical device (IVD)

Lot Number: J59028
ARTG Number 186315
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action At release of the kit lot and at all previous stability testing timepoints, QC specifications were met. However, results obtained at the latest scheduled timepoint (03 Jun 2025) indicated a decreasing bias of the QC panel, which is now at a mean bias of -23% from initial release.

No similar occurrences have been observed within the past two years.
Instructions for Customers/Users Customers are to:
- Quarantine Lot number J59028 immediately
- Complete the Customer Response Form to request replacement product, or to indicate having no affected stock
- Destroy affected stock of Lot number J59028 once replacement product is received.

This action has been finalised on 18/9/2025.
Action Commencement Date 24/06/2025
Responsible Entity Abacus dx Pty Ltd
Contact Information 1800 222 287 - Abacus dx Pty Ltd