| TGA Action ID |
RC-2025-RN-00558-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Universal Joint Screwdriver
Catalogue Number: 6003-100-110
Lot number: 1000530579
GTIN: 04546540502087 |
| ARTG Number |
141514 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The tip component of the Universal Joint Screwdriver may break off during use. The devices utilise a clamping tool that delivered too much force during manufacturing, resulting in product deformation.
There is a potential for breakage to occur during use, which could result in fragments within the surgical field. Fractured metal that enters a patient may require medical intervention.
Stryker will replace any affected product. |
| Instructions for Customers/Users |
Customers are to immediately review the Customer Letter and dispose of any affected product. A Stryker representative will help coordinate the replacement of any affected product. |
| Action Commencement Date |
19/06/2025 |
| Responsible Entity |
|
| Contact Information |
postmarketssp@stryker.com - Stryker |