Print version of this report
TGA Action ID RC-2025-RN-00558-1
Type of Product Medical Device
Product Name/Description Universal Joint Screwdriver

Catalogue Number: 6003-100-110

Lot number: 1000530579

GTIN: 04546540502087
ARTG Number 141514
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action The tip component of the Universal Joint Screwdriver may break off during use. The devices utilise a clamping tool that delivered too much force during manufacturing, resulting in product deformation.

There is a potential for breakage to occur during use, which could result in fragments within the surgical field. Fractured metal that enters a patient may require medical intervention.

Stryker will replace any affected product.
Instructions for Customers/Users Customers are to immediately review the Customer Letter and dispose of any affected product.

A Stryker representative will help coordinate the replacement of any affected product.
Action Commencement Date 19/06/2025
Responsible Entity Stryker Australia Pty Ltd
Contact Information postmarketssp@stryker.com - Stryker