TGA Action ID RC-2025-RN-00556-1
Type of Product Medical Device
Product Name/Description Venclose RF Ablation Catheter

Product codes: VCOUS6F100, VCOUS6F60

Lot numbers: 74507761, 79934996, 74507759, 73573950
ARTG Number 400530
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Specific lots of Venclose RF Ablation Catheters may have been manufactured with internal wiring errors. This error in wire configuration would allow inaccurate temperature detection and may result in devices reaching temperatures higher than intended, while the user receives incorrect warnings that the catheter has not reached temperature.

BD has also received customer complaints which include the following:

Mis-Wiring Problem:
The internal wiring error associated with the Venclose RF Ablation Catheter may result in excessive heat that could cause the following problems, potentially requiring additional medical care should they occur: heat-induced injury to nerves, skin burns, difficulty removing the device, damage to treated vessels or adjacent vessels, endothermal heat-induced thrombosis, deep vein thrombosis, pain, skin discoloration, or embolism of device components. Thrombosis and embolism of device component components could lead to pulmonary embolism.

Battery Power Problem:
Undistributed catheters from lots with potential crossed wires underwent additional inspection to ensure this problem was not present. In some instances, during this inspection, battery tabs became crinkled when reinserted. This may result in catheters arriving nonfunctional due to a lack of battery power. Potential health hazards include a delay to care and discomfort with minor injury.
Instructions for Customers/Users Customers are to:
- Discontinue use of affected product and immediately dispose of product in accordance with the local facility’s processes.
- A credit will be provided once the customer response form is returned.

BD advises that patients who have undergone treatment with a Venclose RF Ablation Catheter that experienced difficulties in achieving the desired temperature may require additional follow-up beyond the usual standard of care at the discretion of the treating physician.

This action has been closed out on 21/07/2026.
Action Commencement Date 18/06/2025
Responsible Entity Bard Australia Pty Ltd
Contact Information 1800 656 100 - Bard Australia