| Reason for the Action |
According to IEC 60601-1:2005+AMD1:2012+AMD2:2020, clause 9.4.2.3 ‘Instability from horizontal and vertical forces’, if an equipment overbalances when being pushed/ leaned on/ etc. with 15% of its weight, the medical electrical equipment shall be marked with a clearly legible warning, e.g., safety sign according to ISO 7010-P017.
Starting April 21, 2021, MCP implemented a safety sign on newly produced INFUSION POLEs but did not correct INFUSION POLEs manufactured before this date. Therefore, all INFUSION POLEs manufactured before April 21, 2021, are affected by this problem.
Given the limited traceability of INFUSION POLEs to SPRINTER CARTs, along with the distribution methods used for the INFUSION POLEs, all SPRINTER CARTs and SPRINTER CARTs XL are considered to be potentially impacted and within the scope of this product correction.
The SPRINTER CARTs and SPRINTER CART XL may overbalance, which could damage the CARDIOHELP/ ROTAFLOW/ ROTAFLOW II or disconnect or damage the perfusion disposable and result in no or low blood flow to the patient. |