TGA Action ID RC-2025-RN-00544-1
Type of Product Medical Device
Product Name/Description VITROS Chemistry Products Performance Verifier I and II. An in vitro diagnostic medical device (IVD).

Used with VITROS Chemistry Products CHE Slides

Catalogue numbers: 806 7324, 823 1474, 191 4605, 800 4707

Lots: All lots within expiry
ARTG Number 232247, 245278
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action QuidelOrtho determined that VITROS Performance Verifiers I and II, when stored at 2–8°C, maintain reconstituted stability for up to 48 hours. However, they may not maintain this stability for up to 7 days for CHE, as specified in the Performance Verifier Assay Sheets.

A drift in the results of VITROS Performance Verifiers I and II below established ranges can lead to QC failure, potentially delaying patient results. In certain cases, this drift in QC results may result in the acceptance of a sub-optimal calibration, which could produce biased patient results. The maximum observed (simulated) bias was noted to be +0.68 U/mL. QuidelOrtho does not recommend reviewing previous results, citing the low likelihood of biased patient outcomes.
Instructions for Customers/Users Customers are to monitor VITROS CHE Slides QC results to identify any drift in VITROS Performance Verifiers I and II falling below 2x Within-lab SD limits. If such a drift is observed, QC should be repeated using freshly reconstituted vials of VITROS Performance Verifiers I and II.
Action Commencement Date 12/06/2025
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd