| TGA Action ID |
RC-2025-RN-00541-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
INVOS Reusable Sensor Cables for PM7100
Part numbers: PMAC71RSC, PMAC71RIC
GTIN numbers: 10884521780125, 10884521806436
Lot numbers: 2023-02-22, 2023-04-28, 2023-05-04, 2023-07-20, 20221220, 20230721 |
| ARTG Number |
304037 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
This recall follows receipt of customer reports of system error messages including “Check Sensor” or “Replace Sensor” displayed on the monitor.
These two error messages can occur during use for numerous reasons by design; however, it was discovered that some reusable sensor cables contain manufacturing defects which can also result in the “Check Sensor” or “Replace Sensor” error messages.
Corrective actions have been implemented at the manufacturing facility to resolve these problems going forward. |
| Instructions for Customers/Users |
Quarantine and stop using all product from the affected lot of INVOS Reusable Sensor CablePMAC71RSC and PMAC71RIC. Affected product should be returned to Medtronic.
Medtronic will proactively send free replacement cables to those customers with impacted product (based on prior shipments) at the earliest product availability.
This action has been finalised on 22/9/2025. |
| Action Commencement Date |
11/06/2025 |
| Responsible Entity |
|
| Contact Information |
0402 897410 - James Woodin, Medtronic Australasia |