TGA Action ID RC-2025-RN-00541-1
Type of Product Medical Device
Product Name/Description INVOS Reusable Sensor Cables for PM7100

Part numbers: PMAC71RSC, PMAC71RIC

GTIN numbers: 10884521780125, 10884521806436

Lot numbers: 2023-02-22, 2023-04-28, 2023-05-04, 2023-07-20, 20221220, 20230721
ARTG Number 304037
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action This recall follows receipt of customer reports of system error messages including “Check Sensor” or “Replace Sensor” displayed on the monitor.

These two error messages can occur during use for numerous reasons by design; however, it was discovered that some reusable sensor cables contain manufacturing defects which can also result in the “Check Sensor” or “Replace Sensor” error messages.

Corrective actions have been implemented at the manufacturing facility to resolve these problems going forward.
Instructions for Customers/Users Quarantine and stop using all product from the affected lot of INVOS Reusable Sensor CablePMAC71RSC and PMAC71RIC. Affected product should be returned to Medtronic.

Medtronic will proactively send free replacement cables to those customers with impacted product (based on prior shipments) at the earliest product availability.

This action has been finalised on 22/9/2025.
Action Commencement Date 11/06/2025
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information 0402 897410 - James Woodin, Medtronic Australasia