TGA Action ID RC-2025-RN-00534-1
Type of Product Medical Device
Product Name/Description ABL90 FLEX and ABL90 FLEX PLUS analysers with software version 3.5 MR11 and below

Product codes: 393-090, 393-092
ARTG Number 228382
Action Type Product Alert
Action Level Hospital
Hazard Classification Class III
Reason for the Action Upon restarts the analyser may fail to initialise the calibration schedule, which means that the sensors will not be calibrated. The calibration schedule will most likely be initialised by restarting the analyser.

The default setup for internal quality control measurements will flag out-of-specification sensor performance. The described error does not impact analyser performance. Therefore, no risk for the patient.
Instructions for Customers/Users Customers are to be aware of the problem outlined in the customer letter. The described error does not impact analyser performance and can continue to be used.

For analysers running with Windows 10:
Radiometer is developing an updated analyser software that will eliminate the potential problem.

For analysers running with Windows XPE or Windows 7:
Radiometer is working to identify a solution for these analysers.

A separate market action will be conducted when the software corrections are available.

This action has been closed out on 03/07/2026
Action Commencement Date 12/06/2025
Responsible Entity Radiometer Pacific Pty Ltd
Contact Information 1800 247 254 - Radiometer Pacific