Print version of this report
TGA Action ID RC-2025-RN-00529-1
Type of Product Medical Device
Product Name/Description Zenition 70 Systems

Model Number: 718133
ARTG Number 405770
Action Type Product Correction
Action Level Hospital
Hazard Classification Class III
Reason for the Action Philips has identified that, when the Wireless Foot Switch is used for vascular procedures (i.e., with the vascular protocol) in conjunction with the Zenition 70 system, there may be a delay of up to 72 milliseconds in the command to stop the radiation when the user releases the pedal of the Wireless Foot Switch. This problem can only occur during a vascular procedure using 2fps (frames per second) subtract mode. During a typical vascular procedure, the vascular protocol may be used up to 15 times, resulting in a maximum estimated additional radiation dose to the patient of 0.0012 mSv.

Based on this minimal additional dose, it is considered unlikely to contribute to any immediate or long-term health effects from radiation exposure. To date, Philips has not received any complaints associated with this problem.
Instructions for Customers/Users Customers are to note that Philips will install a software update to address this problem (reference FCO71800104). Beginning in June 2025, a Philips representative will contact customers to arrange a visit for the installation of the update.

Update 3/10/2025: The corrections have been released. Philips have commenced contacting customers to schedule on-site visits for the installation of these corrections

This action has been closed out on 23/04/2026
Action Commencement Date 6/06/2025
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team