| TGA Action ID |
RC-2025-RN-00528-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Philips Zenition 50, Zenition 70
Zenition 50 Product Code: 718096 UDI: (01)00884838091535(21)
Zenition 70 Product Code: 718133 UDI: (01)00884838091528(21) |
| ARTG Number |
405770 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
A wireless foot switch pedal may get stuck in the active position when the user releases the pedal, resulting in the emission of unintended radiation because of:
· Build-up of dense or sticky fluids on the wireless foot switch (e.g., blood or contrast fluid), if not properly cleaned. · Use of protective covers that are either the wrong size or incorrectly placed on the wireless foot switch. · Dislodgment of a screw holding the pick-up bar of the wireless foot switch, subsequently becoming lodged within the wireless foot switch housing. · An additional screw inadvertently left inside the wireless foot switch housing during manufacturing.
To date, Philips has not received any complaints reporting unintended radiation due to the wireless foot switch pedal getting stuck. |
| Instructions for Customers/Users |
Philips will contact customer to schedule a visit to inspect the wireless foot switch and hand over an IFU addendum (reference FCO71800111).
This action has been closed out on 19/06/2026 |
| Action Commencement Date |
5/06/2025 |
| Responsible Entity |
|
| Contact Information |
1800 251 400 - Philips Service Delivery Team |