TGA Action ID RC-2025-RN-00510-1
Type of Product Medical Device
Product Name/Description VentrEX Complete EVD System C

Product Code: NRG-58-00010C

Lot Numbers: 410167, 411060, 407140, 501022, 412145
ARTG Number 410831
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Life Healthcare have identified that a warning notice was not included for specific lot numbers of VentrEX Complete EVD System 3-way stopcock.

The proximal 3-way stopcock of the patient line in the VentrEX Complete EVD System REF 58-00010C is permanently affixed to the disposable stand. This may lead to an increased risk under unfavourable circumstances:

*If the disposable stand is screwed onto a drip stand and the user then opens and closes (i.e., rotates) the handle of this 3-way stopcock, higher tensile and lever forces are generated than with a stopcock that is only snap-fitted into the stand (standard version).
* If the user applies tensile or lever forces while turning the handle (through a twisting pull motion), the surrounding spring of the handle can be damaged and the handle can be either fully withdrawn or partial loosened, potentially leading to uncontrolled drainage or system leakage.
Instructions for Customers/Users Customers should check their internal inventory and retain a copy of the customer letter with the missing warning for their records.

This action has been closed out on 21/07/2026
Action Commencement Date 3/06/2025
Responsible Entity Life Healthcare Pty Ltd