TGA Action ID RC-2025-RN-00509-1
Type of Product Medical Device
Product Name/Description Various Texium Components

Product Names: BD Texium 3 mL, 5ml, 10ml, 20ml, 60ml & BD Texium closed-male luer with female cap

Product Codes:MY8003-0006, MY8005-0006, MY8010-0006, MY8020-0006, MY8060-0006, 10012241

Multiple lot numbers

ARTG's: 177502, 307702
ARTG Number 177502, 307702
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action BD has identified that specific lots of standalone Texium components, and IV infusion sets and syringes containing the Texium closed-male Luer component (as listed above) have the potential to leak during use at the connection interface between the Texium component and the SmartSite Needle-free Connector.
Instructions for Customers/Users Customers should;
1. Replace affected products in use.
2. If already in use, and you are unable to determine if the product is within scope, continue to monitor for leakage or replace product.
3. BD will provide replacement/credit for all discarded inventory

This action has been finalised on 22/9/2025.
Action Commencement Date 3/06/2025
Responsible Entity Becton Dickinson Pty Ltd
Contact Information 1800 656 100 - Customer Service Team