| TGA Action ID |
RC-2025-RN-00505-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
HeartSine samaritan PAD
Model Numbers: 350P, 360P, 500P
Product Codes: 350BASUK10, 360BASUK10, 360BASUKGW, 500BASUK10
A subset of serial numbers beginning with 21, 22, 23 or 24, and are then followed by the letter B, D, E, G or H
ARTG's: 156690 & 228210 |
| ARTG Number |
156690, 228210 |
| Action Type |
Recall |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
It was determined during internal quality testing that a manufacturing process problem related to a circuit board component may impair the device’s ability to function or cause failure. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. The device becomes inoperable after the failure occurs. If this problem occurs, the device may fail to deliver the intended therapy during use, potentially leading to a delay in treatment or no treatment being delivered during use. The problem was observed during quality testing, not during patient use. There have been no in-field product complaints or adverse events reported related to this product problem.
Update 21/7/2026: 2 New Problems have been identified 1: A problem with the membrane tail, an internal component, which may result in an intermittent fault. This failure could occur at any point when the device is holding a charge in preparation to deliver therapy, while delivering a shock, or after shock delivery. If this failure occurs, the device becomes inoperable.
2: A limited number of HeartSine Samaritan PAD devices were distributed without the required maintenance supplement. It contains periodic maintenance instructions implemented as an additional risk control measure. |
| Instructions for Customers/Users |
To identify affected serial numbers follow the link: https://www.stryker.com/us/en/emergency-care/product-notices/heartsine/index.html
DISTRIBUTORS: - Inspect inventory to identify affected stock and contact Stryker to arrange for replacement stock.
- If any affected devices have been sold to end customers notify Stryker immediately and Stryker will contact these customers.
CONSUMERS: - Inspect inventory to identify affected stock and contact Stryker to arrange for replacement stock. Stryker does not currently have unaffected stock available. Stryker are working closely with the manufacturer and will provide an updated timeline once available.
In the interim, Stryker recommends keeping HeartSine samaritan PADs in service, and maintain awareness of the communication internally until all required actions have been completed.
Update 21/7/2026: Download a copy of the missing maintenance supplement when acknowledging the recall & product correction store with your device’s user manual until a replacement is received. It is recommended that the supplement is stored with your user manual/user handbook and AED and follow the additional check once every 3 months. |
| Action Commencement Date |
20/06/2025 |
| Responsible Entity |
|
| Contact Information |
heartsine.recall@stryker.com - Stryker Australia |