TGA Action ID RC-2025-RN-00500-1
Type of Product Medical Device
Product Name/Description Incisive CT and CT 5300

Model Numbers: 728143, 728144, 728285

Serial Numbers: 500408, 550187, 550241, 800028

UDI: (01)00884838105508(21), (01)00884838085015(21), (01)00884838113237(21)
ARTG Number 352131
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The HIB-CIB cable on the NOAH couch may wear unexpectedly over time. During vertical couch motion, if the HIB-CIB cable has been damaged, the system may engage the Emergency Stop intermittently. If the Emergency Stop is engaged both motion and X-rays will stop.

If this occurs during a scan it could result in a partial set of images, and a patient rescan may be required.
Instructions for Customers/Users Customers may continue to use the Philips CT system(s) in accordance with the intended use.

If customers experience the Emergency Stop opening intermittently during vertical couch motion and cannot be solved by restarting, they should contact the Philips Service Delivery Team on 1800 251 400 for further support.

Philips will contact customers to schedule a time for a Philips Field Service Engineer (FSE) to visit the site and install the solution to resolve this problem.

This action has been closed out on 30/03/2026
Action Commencement Date 28/05/2025
Responsible Entity Philips Electronics Australia Ltd
Contact Information 1800 251 400 - Philips Service Delivery Team