TGA Action ID RC-2025-RN-00478-1
Type of Product Medical Device
Product Name/Description Ceiling Mounted Tubestand

Model Number: CH-200 and CH-200M

Serial number: All
ARTG Number 106713
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The ceiling-mounted X-ray tube suspension systems, CH-200M and CH-200M, consist of a ceiling suspension mechanism and an X-ray tube holder (hereafter referred to as "holder"), to which the X-ray tube is securely attached.

The design includes a rotation mechanism that allows for adjustment of the X-ray tube to achieve optimal beam alignment. Additionally, there are protective components installed to prevent the holder from detaching, even in the event of a retaining shaft failure.

However, based on evaluations and statistical data from operational use, there is a low-risk potential that these parts may not function effectively if the retaining shaft fails during operation.
Instructions for Customers/Users · Customers are to perform daily inspections in accordance with the CH -200/CH -200M operation manual.
· Customers are to check specifically for any abnormalities such as the X-ray tube device rotating around the vertical axis without pressing any control panel buttons.
· Customers are to stop using the equipment immediately and contact Shimadzu service representative if any abnormal points are observed.
· Customers are to avoid rotating the X-ray tube assembly above or near the patient.
· Customers are aware that Shimadzu will implement corrective measures by adding additional parts to prevent the X-ray tube holder from coming off in the event of a retaining shaft failure.
· Customers are aware that the corrective measures are planned for completion within 12 months.

This action has been closed out on 11/05/2026
Action Commencement Date 19/05/2025
Responsible Entity Shimadzu Oceania Pty Ltd
Contact Information 02 9898 2444 - Shimadzu Oceania Pty Ltd Customer Service