| TGA Action ID |
RC-2025-RN-00476-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Qube Nitrile Examination Gloves
Size: Large
Batch number: 20438
Manufacture date: 1-Aug-2022
Expiry date: 31-July-2027 |
| ARTG Number |
363826 |
| Action Type |
Recall |
| Action Level |
Wholesale |
| Hazard Classification |
Class II |
| Reason for the Action |
TGA testing found that the affected batch did not meet the required specifications for Freedom from Holes (Watertightness). A non-compliant result was also reported by independent testing. This deviation from expected standards may impact the product’s quality and efficacy.
An investigation is underway to determine the root cause. Almitas Group is recalling the affected products. |
| Instructions for Customers/Users |
Customers are to: - Quarantine then return affected stock on hand. - Contact enquiries@almitasgroup.com to arrange for replacement stock. |