| TGA Action ID |
RC-2025-RN-00454-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Urea. An in vitro diagnostic medical device (IVD).
Catalogue Number: UR446
Batch/Lot: 606771
UDI: 05055273206920
Expiry Date: 28/07/2025 |
| ARTG Number |
199617 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The R1 b Enzyme Reagent is not meeting its stability claims once reconstituted and is not stable for 4 weeks at +2 to +8°C or 2 days at +15 to +25°C. The calibration factor, QC's and patient samples decrease during repeat testing by up to -5% within 3hrs of reagent reconstitution.
Affected batch has the potential to misclassify patient samples as results will be falsely decreased depending on testing time from reagent reconstitution. |
| Instructions for Customers/Users |
Customers are to: - Discontinue use of and discard any of the affected batch. - Provide Randox with photographic evidence of the destruction of the kits for reimbursement. - Review results generated with the affected batches in line with the clinical profile of the patient.
Randox Laboratories will replace any discarded inventory.
This action has been closed out on 31/07/2026 |
| Action Commencement Date |
8/05/2025 |
| Responsible Entity |
|
| Contact Information |
02 9615 4640 - Randox Australia Office |