| TGA Action ID |
RC-2025-RN-00447-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Clinician Programmer
Origin Data Management (ODM) software versions 3.1.0, 3.1.1, 3.1.2, 3.2.0, 3.2.1,
UDI-DI: 04056481145118 (ODM v. 3.1) & 04056481146078 (ODM v. 3.2). |
| ARTG Number |
308574 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Boston Scientific is informing customers that under certain conditions, patient records from unrelated individuals may be automatically merged into a single record without notifying the user. |
| Instructions for Customers/Users |
Customers should read the instructions in the customer letter for further information on how this may occur and how to remedy the problem.
Before patient treatment, always: · Verify that the name of the data set and/or treatment plan and its date are correct. · Verify that the images and/or plan are correct for the intended treatment.
Brainlab will provide a software solution, in which the described problem is corrected, to all affected customers as early as May 2025 to schedule the update.
This action has been closed out on 03/06/2026 |
| Action Commencement Date |
8/05/2025 |
| Responsible Entity |
|
| Contact Information |
ANZ_Incident_Report@bsci.com - Boston Scientific |