Print version of this report
TGA Action ID RC-2025-RN-00443-1
Type of Product Medical Device
Product Name/Description Euky Bear (Warm Steam) Vaporiser

Model EBSV2013

All batches
ARTG Number 297553
Unique Device Identifier (UDI) n/a
Action Type Recall
Action Level Consumer
Hazard Classification Class II
Reason for the Action This communication follows the 2023 Recall RC-2023-RN-00777-1.

The enclosed heating element for these batches is overheating and / or the polymer may disintegrate, if the device is run dry. This may result in the device producing smoke or fumes. If the heating module finger is exposed, it could cause burns.

This recall does not affect any other model of Euky Bear (Warm Steam) Vaporiser other than model EBSV2013 or any other Felton Grimwade & Bosisto’s products.
Instructions for Customers/Users Consumers are to:

-Immediately stop using Euky Bear (Warm Steam) Vaporiser Model number EBSV2013, unplug it, cut the power cord and dispose of it.

-Email eukybear@fgb.com.au to arrange a replacement.

The affected model number, EBSV2013, can be found on the sticker on top of the blue heating module near the power cord.

-Retailers are to refer consumers to FGB and guide them on identification of model number.

The affected model number has been superseded and has not been supplied to customers since February 2023.

This action has been closed out on 19/08/2025
Action Commencement Date 19/05/2025
Responsible Entity Felton Grimwade & Bosisto's Pty Ltd
Contact Information 1800 655 841 - FeltonGrimwade and Bositos