Print version of this report
TGA Action ID RC-2025-RN-00439-1
Type of Product Medical Device
Product Name/Description Medtronic Evolut Bioprosthesis

Multiple products and model numbers

ARTG's 284003, 319850, 430190
ARTG Number 284003, 319850, 430190
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Medtronic has received reports of users having difficulty opening the jar, which may cause damage to the lid, resulting in the potential for particulate transferring from the lid to the jar containing the bioprosthesis.

As of March 31, 2025, Medtronic has observed an incidence rate of 0.125% for this jar lid problem in Australia and New Zealand. To date, there has been no patient harm nor serious injury reported.

Medtronic is conducting a thorough investigation of the events, including manufacturing process mitigations.

In the interim, Medtronic has initiated an action to update the applicable IFUs for Evolut FX as follows:
“Do not use the bioprosthesis if there is any damage to the container (for example, cracked jar or lid, leakage, particulate material, broken or missing seals or jar lid gasket).”
Instructions for Customers/Users Medtronic has no evidence of seal degradation and no evidence that the sterility of the bioprosthesis is compromised. If particulate material is observed in the solution upon opening the jar, the bioprosthesis should be discarded and a new unopened device used instead.

In accordance with current Instructions for Use (IFU) and training, the bioprosthesis should always be rinsed with 3 rinsing bowls before use.

Customers should be aware of the IFU update for Evolut FX as follows:
“Do not use the bioprosthesis if there is any damage to the container".

Medtronic does not recommend any further actions for patients already treated with the listed devices and patients should be managed according to the standard of care after a TAVI procedure.

This action has been finalised on 27/5/2026.
Action Commencement Date 5/05/2025
Responsible Entity Medtronic Australasia Pty Ltd
Contact Information 0429120212 - katina.weston@medtronic.com - Katina Weston