| Reason for the Action |
Medtronic has received reports of users having difficulty opening the jar, which may cause damage to the lid, resulting in the potential for particulate transferring from the lid to the jar containing the bioprosthesis.
As of March 31, 2025, Medtronic has observed an incidence rate of 0.125% for this jar lid problem in Australia and New Zealand. To date, there has been no patient harm nor serious injury reported.
Medtronic is conducting a thorough investigation of the events, including manufacturing process mitigations.
In the interim, Medtronic has initiated an action to update the applicable IFUs for Evolut FX as follows: “Do not use the bioprosthesis if there is any damage to the container (for example, cracked jar or lid, leakage, particulate material, broken or missing seals or jar lid gasket).” |