TGA Action ID RC-2025-RN-00437-1
Type of Product Medical Device
Product Name/Description Odontopaste OP8G

Batch Numbers: B3066, B3066F, B3066H

Product Code: OP8G

Expiry Date: Mar2026
ARTG Number 148898
Action Type Recall
Action Level Hospital
Hazard Classification Class III
Reason for the Action Products were stored outside of the defined parameters of 5-25ºC. A short-term allowance of up to 2 weeks at temperatures up to 40ºC is supported by validation studies.

There is a risk that the efficacy of the device could be negatively impacted and/or unintended breakdown products could be formed.

However, no customer complaints, injuries, or adverse events related to this problem have been reported.
Instructions for Customers/Users Customers are to:
1.Immediately identify all affected lot numbers and quarantine them on-site to prevent further use.
2.Await collection arrangements, as the sponsor will coordinate the retrieval of all affected products.
3.Expect re-supply of replacement items from the distributor for all affected material that are returned.

This action has been finalised on 23/9/2025.
Action Commencement Date 5/05/2025
Responsible Entity M & A Athanassiadis Dental Pty Ltd T/A Australian Dental Manufacturing
Contact Information 07 3878 1901 - Alex Athanassiadis