| TGA Action ID |
RC-2025-RN-00425-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Hugo RAS (Robotic-Assisted Surgery)
Product Identifiers: C22CAB0087, C22CAB0088, C22CAJ0148
GTIN: 10884521826663, 10884521836266
Model Number: MRASC0005 |
| ARTG Number |
377742 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
An excessive torque error is occurring for instruments at an increased occurrence rate. This error prevents further use of the instrument and requires it to be removed. As a result of the disabled instrument, it has been observed that users may place excessive torque on a robotic arm when attempting to remove a disabled instrument. This incorrect approach to instrument removal may activate a safety mechanism that disables the robotic arm and requires the robotic arm to be rebooted before further use.
As of March 11, 2025, Medtronic has received 290 complaints related to this matter. Nine reports mention extended procedure durations over 30 minutes, three involve alternative approaches leading to potential tissue damage or trauma, and one report involves bleeding.
This action does not impact patients who have previously undergone a procedure with the Hugo RAS system. |
| Instructions for Customers/Users |
- Medtronic is updating the labelling to clarify the necessary steps to release disabled instruments without unnecessarily triggering the system's safe mode. This update will be included in the next revision of the User Guide. -Medtronic is providing supplemental labelling in the Customer Letter to be used until the User Guide is updated. - An enhanced version of the Hugo RAS system software has been developed to reduce the occurrence of excessive torque errors. - A Medtronic service representative will assess the current software version installed on the Hugo RAS system and, if necessary, will install the enhanced version to affected customers.
This action has been finalised on 27/10/2026. |
| Action Commencement Date |
29/04/2025 |
| Responsible Entity |
|
| Contact Information |
0407 464 883 - rs.anzpmv@medtronic.com - Medtronic Australasia |