TGA Action ID RC-2025-RN-00424-1
Type of Product Medical Device
Product Name/Description ORTHO VISION Analyzer and ORTHO VISION Max Analyzer. An in vitro diagnostic medical device (IVD).

Unique Device Identifier: 6904579, 6904578

Affected Software: Version 5.16 and below

Lot: All current lots
ARTG Number 343753
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The ORTHO VISION Analyzer and ORTHO VISION Max Analyzer, which are used for BioVue Cassettes, were found to have a software anomaly. This problem occurs when the analyzer is unable to perform an inventory of loaded samples and reagents if the loading station door is opened immediately after the allotted 20-second countdown and within the 0.2 second window before the door re-locks.

Failure to complete sample re-inventory in this scenario may lead to the mis-association of samples on the tray if new samples are loaded, removed or samples are moved to different locations, potentially resulting in erroneous patient results.
Instructions for Customers/Users • Customers are to visually confirm that the timer has not reached the end of its 20 second window before accessing the loading station door. If the timer has expired, they must allow the door to reengage before accessing the load station through the user interface.
• Customers are to initiate a new loading event if the loading door was opened at the '0:00' time mark and no inventory is detected on the GUI. To do so, they should open the loading station door, wait a few seconds and then close it to ensure an accurate inventory event is facilitated.

Update 11/05/26: Software version (APSW) 5.17.0, Modification (MOD) 96, and Operating System (OS) 4.6.0 (MOD 94) for ORTHO VISION and ORTHO VISION Max (Swift) Analyzers are now available to resolve this problem.

Customers are to:
- Install software version (APSW) 5.17.0, Modification (MOD) 96, and Operating System (OS) 4.6.0 (MOD 94).

For further information on the installation process, see the customer letter.


This action has been closed out on 28/08/2026
Action Commencement Date 30/04/2025
Responsible Entity Ortho-Clinical Diagnostics Australia Pty Ltd
Contact Information 1800 032 359 - Technical Services Centre