Print version of this report
TGA Action ID RC-2025-RN-00420-1
Type of Product Medical Device
Product Name/Description DxI 9000 Access Immunoassay Analyzer. An in vitro diagnostic medical device (IVD).

REF: C11137

Software Version: 1.20
ARTG Number 177999
Unique Device Identifier (UDI) 15099590732103
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Beckman Coulter has determined that DxI 9000 Access Immunoassay Analyzers on software version 1.20 may experience an increase in reaction vessel (RV) supply homing failures during system initialisation.

These homing failures will cause the analyser to fail initialisation and will result in a red system event with the following description: “RVSupplyAssembly – MachineControlStatus (RVSupply was not homed)”.

If repeated initialisation attempts continue to fail and a service visit is required to resolve the problem, there may be a delay in reporting test results which could lead to a delay in patient treatment.

This problem does not cause erroneous patient test results.
Instructions for Customers/Users If your laboratory has experienced the red system event and it is not resolved by initialising the analyser, contact your Beckman Coulter representative and schedule a service visit.

A Beckman Coulter representative will also schedule the installation modification when it is available.

This action has been finalised on 28/7/2025.
Action Commencement Date 24/04/2025
Responsible Entity Beckman Coulter Australia Pty Ltd
Contact Information 02 9844 6000 - Sara Panahi