TGA Action ID RC-2025-RN-00415-1
Type of Product Medical Device
Product Name/Description 3M Ranger Blood/Fluid Warming High Flow Set and 3M Ranger Fluid Warming Irrigation Set

Catalog Numbers: 24355, 24750

Lot numbers: all lot numbers manufactured after March 2022

ARTG's 219153, 219148
ARTG Number 219148, 219153
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Through an investigation by the product support team, it was determined the design requirements did not clearly specify the location of the temperature measurement and corresponding flow rate based on inlet temperature. As a result, it is recommended the design specifications and labelling be clarified for these items.

If the device is used at a flow rate outside of the updated information provided, there is a potential risk for the device to administer the fluids at a lower temperature than indicated and cause a decrease in patient temperature.

To date, there have been no hypothermia-related complaints reported associated with the use of the device.
Instructions for Customers/Users Customers may continue to use the products as indicated, taking into consideration the updated information provided below and in the customer letter.

Design requirements will be updated to specify the location of the temperature measurement and corresponding flow rate based on inlet temperature and associated documents:

• Hardware manual
• IFU documentation
• Design control matrix

· Keep the notice on hand until you are notified that the updated IFUs and operator’s manuals are available.

This action has been finalised on 16/3/2026.
Action Commencement Date 1/05/2025
Responsible Entity KCI Medical Australia Pty Ltd
Contact Information 1300 524 822 or anzfieldaction@solventum.com