TGA Action ID RC-2025-RN-00409-1
Type of Product Medical Device
Product Name/Description Elastomeric Infusion Pump System

Multiple product descriptions and product codes
ARTG Number 168893
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Baxter Healthcare has received reports of under-infusion involving compounded medicines administered via elastomeric infusion devices.
This problem may result in patients receiving less than the prescribed dose of their therapy within the intended infusion timeframe.
Some reports specifically concern a combination of medications used in the EPOCH oncology regimen - Doxorubicin, Etoposide, and Vincristine - which are compounded upon request under an Elastomeric Stability Agreement.
Several factors can influence the flow rate of elastomeric infusion devices, including:
(1) Fill Volume, (2) Temperature, (3) Viscosity of the medication, (4) The difference in height between the Fill Port and the Distal End Luer Lock/Flow Restrictor and (5) Catheter size.

Update 29/09/2025: An update to the IFU to include additional information on factors that may impact flow rate is in-progress. This will be available in 2026.

Meanwhile, Baxter has implemented the following measures to support customers with additional guidance on factors that may impact flow rate:
·Updated Elastomeric Products Professional Guide
·Updated Elastomeric Patient Guide and Elastomeric Mini Patient User Guide.
Instructions for Customers/Users Update 29/09/2025:
- Customers are to be aware of the resources available electronically via Baxter Professional on-line portal and Baxter Compounding App which is free to access for HCPs.

- Patients can also access Elastomeric Patient Resources | Home Patient Resource Portal.

This action has been finalised on 9/7/2026.
Action Commencement Date 28/04/2025
Responsible Entity Baxter Healthcare Pty Ltd
Contact Information 02 9848 1111 - Baxter OneCall Medical Information