| TGA Action ID |
RC-2025-RN-00401-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
FRED easyport
Model Number: SCH-0.900000
Devices with IGBT module manufactured in 2020 and onwards.
Multiple serial numbers |
| ARTG Number |
174306 |
| Action Type |
Recall |
| Action Level |
Consumer |
| Hazard Classification |
Class II |
| Reason for the Action |
SCHILLER AG has identified isolated cases where the FRED easyport failed routine maintenance tests due to defective IGBT modules. Failure of the IGBT module may result in the device failing the maintenance by being unable to deliver a shock. However, it may be possible to deliver a limited number of shocks before failing to deliver further shocks.
If regular maintenance is not carried out at the intervals specified in the IFU, a defective IGBT module may go undetected. As a result, excessive leakage current in the module could prevent the device from fully charging the capacitor, thereby limiting the number of shocks it can deliver before failing to deliver further shocks.
This problem has occurred exclusively during maintenance, and no incidents involving patients have been reported.
All known cases of the defect involve devices equipped with IGBT modules manufactured in 2020 or later. |
| Instructions for Customers/Users |
Customers are to refer to appendix A in the customer letter to identify if they have an impacted device with an IGBT module manufactured in 2020 and onwards.
Please note the FRED easyport is no longer manufactured, so the affected device will be replaced with FRED easyport plus. Once the replacement device is received by customers, return the impacted device to InterMed.
In the meantime, customers may continue using the existing FRED easyport device, provided that all maintenance requirements outlined in the customer letter are strictly followed.
All users are requested to read the current valid Instructions for Use (attached to the customer letter) and take note of the lifetime in chapter 7.2 Safety standard.
This action has been finalised on 23/6/2026. |
| Action Commencement Date |
11/07/2025 |
| Responsible Entity |
|
| Contact Information |
1300 123 633 - Regulatory Affairs & Quality Assurance Manager ANZ |