TGA Action ID RC-2025-RN-00395-1
Type of Product Medical Device
Product Name/Description TufTex Over-the-Wire Embolectomy Catheter, Pruitt Occlusion Catheter & Pruitt Irrigation Occlusion Catheter

TufTex Over-the-Wire Embolectomy Catheter
Item Numbers: 1651-38, 1651-44, 1651-48, 1651-68, 1651-78, 1651-84, 1651-88

Pruitt Occlusion Catheter
Item Numbers: 2103-36, 2103-46, 2103-56

Pruitt Irrigation Occlusion Catheter:
Item Number: 2102-09

ARTG's: 147757 & 214149
ARTG Number 147757, 214149
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action During product testing packages were found to have incomplete seals (sterile barrier).

Some parts (approx. 10-15%) could exhibit voids on the seal area where the tubing connects to the boat tray. The voids were observed visually against backlight and confirmed upon opening of Tyvek lid. Seal integrity testing of other samples with visually detected voids confirmed failures in some of the samples, indicating compromised packaging.
Instructions for Customers/Users Customers are to:
- Identify and quarantine any impacted product from your inventory.
- Contact the sponsor indicating number of affected devices and request an RMA number to return the products.
- Return all impacted products to the sponsor.

This action has been finalised on 2/10/2025.
Action Commencement Date 15/04/2025
Responsible Entity Emergo Asia Pacific Pty Ltd T/a Emergo Australia
Contact Information 02 9006 1662 - Emergo Australia