TGA Action ID RC-2025-RN-00393-1
Type of Product Medical Device
Product Name/Description CADD-Solis and CADD-Solis VIP Ambulatory Infusion Pumps USO Alarm

Multiple product codes and lot numbers
ARTG Number 396633
Unique Device Identifier (UDI) CADD Pump
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Under certain conditions, a CADD-Solis pump may trigger an erroneous (false) Upstream Occlusion (USO) Alarm. It may occur when there is a delay of more than one hour between the first prime or infusion of a new CADD Administration Set and the next prime or infusion of the same CADD Administration Set.

The USO Alarm is a high priority alarm that will interrupt an ongoing infusion or delay initiation of an infusion. The pump cannot resume or start an infusion until the alarm is cleared.

A USO Alarm may potentially occur if all these conditions are met:
• Using a CADD administration set (i.e. not a medication cassette reservoir), and
• Enabling the Upstream Occlusion Alarm, and
• Not programming Keep Vein Open (KVO) or Continuous rate, and
• Priming or infusing shortly after attaching the administration set and the next infusion does not occur for approximately one hour or longer.

This problem will not occur when using a medication cassette reservoir, if the USO Alarm is disabled, if the KVO setting is programmed, or if a continuous rate setting is programmed.

To date, Smiths Medical has not received any reports of serious injuries or deaths related to this problem.
Instructions for Customers/Users Customers are to be aware that a USO alarm may occur when all the conditions stated are met.

The USO alarm can be cleared during these conditions by removing the administration set from the pump. After re-attaching the administration set, delivery can be restarted without an alarm.

The Instructions of Use will be updated to include this information.
Action Commencement Date 17/04/2025
Responsible Entity Smiths Medical Australasia Pty Ltd