| Reason for the Action |
Under certain conditions, a CADD-Solis pump may trigger an erroneous (false) Upstream Occlusion (USO) Alarm. It may occur when there is a delay of more than one hour between the first prime or infusion of a new CADD Administration Set and the next prime or infusion of the same CADD Administration Set.
The USO Alarm is a high priority alarm that will interrupt an ongoing infusion or delay initiation of an infusion. The pump cannot resume or start an infusion until the alarm is cleared.
A USO Alarm may potentially occur if all these conditions are met: • Using a CADD administration set (i.e. not a medication cassette reservoir), and • Enabling the Upstream Occlusion Alarm, and • Not programming Keep Vein Open (KVO) or Continuous rate, and • Priming or infusing shortly after attaching the administration set and the next infusion does not occur for approximately one hour or longer.
This problem will not occur when using a medication cassette reservoir, if the USO Alarm is disabled, if the KVO setting is programmed, or if a continuous rate setting is programmed.
To date, Smiths Medical has not received any reports of serious injuries or deaths related to this problem. |