Print version of this report
TGA Action ID RC-2025-RN-00379-1
Type of Product Medical Device
Product Name/Description Micro-X Rover Mobile X-Ray System

Model: MXU-RV35

Serial Numbers: #335, #363

The Serial Numbers affected have been distributed since November 2023
ARTG Number 374584
Unique Device Identifier (UDI) Please refer to product traceability list attached.
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Micro-X has received several product complaints from customers in the United States regarding failures in the internal fastening mechanism of the generator. These failures have led to the partial separation of the capacitor bank (an integral part of the generator) from the main assembly, rendering the device mechanically inoperable.

Potential risk to the patient is delay in diagnosis and treatment.
There is no increased risk for the operator.
Instructions for Customers/Users Customers are to:

- Quarantine affected stock on hand to prevent further use until a suitable correction is implemented by a Micro-X service agent.

- Contact their nominated Service representative to arrange product correction or Email: support@micro-x.com.

Product correction consists of installing generator capacitor bank support brace to provide suitable generator support.

This action has been closed out on 09/07/2025
Action Commencement Date 8/04/2025
Responsible Entity Micro-X Ltd
Contact Information 08 7099 3966 or admin@micro-x.com - Micro-X