TGA Action ID RC-2025-RN-00377-1
Type of Product Medical Device
Product Name/Description The BioFire Blood Culture Identification 2 (BCID2) Panel. An in vitro diagnostic medical device (IVD).

REF: RFIT-ASY-0147 (30-pack)

All lot numbers

All unexpired products
ARTG Number 224167
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Following the market actions RC-2023-RN-00943-1 and RC-2024-RN-00110-1, BioMérieux Australia is issuing a Product Correction regarding a mandatory pouch module software update. , This software update will mitigate risks linked to the previous market actions by modifying the analysis parameters (RC-2023-RN-00943-1) and adding an annotation to the customer report (RC-2024-RN-00110-1).

False negative results for blood stream infections may cause delays in accurate results. A false positive result of Candida tropicalis could lead to a patient remaining on inappropriate therapy potentially for three or more days, which could be life-threatening.

In addition to the pouch module software update, the BIOFIRE BCID2 Panel Instructions for Use (IFU) have been updated.
Instructions for Customers/Users Customers who are affected will receive a notification from BioMérieux Australia regarding the requirement to update the pouch module software for the BIOFIRE BDID2 panel.
Action Commencement Date 4/04/2025
Responsible Entity Biomerieux Australia Pty Ltd
Contact Information 1800 333 421 - Biomerieux Australia Helpdesk