Print version of this report
TGA Action ID RC-2025-RN-00374-1
Type of Product Medical Device
Product Name/Description Human Assayed Multi-Sera Level 3. An in vitro diagnostic medical device (IVD)

Catalogue number: HE1532, HS2611

Batch number: 1248UE, 1264UE, 1265UE, 1264UE, Packed into batches 610445, 606991

Expiry date: 28th Jan 2026, 28th Dec 2025

Manufacturing date: 2nd May 2022, 19th July 2022, 20th Oct 2022, 06th July 2022
ARTG Number 207886
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Certain lots and batches of the Human Assayed Multi-Sera Level 3 have the analyte NEFA increased over the shelf life of the product and requires the targets and ranges to be reassigned.

As a result, there may be delay in testing patient samples and/or a delay in reporting patient results. However, NEFA is not a standalone assay used for diagnostic testing of the patient’s lipid profile.
Instructions for Customers/Users Randox Laboratories has reassigned the Human Assayed Multi-Sera Level 3 targets and ranges.

The value sheets have been updated and are available on www.randox.com. Customers should discard the previous version and download the updated version.

Randox will provide updates once the root cause has been identified, and the Corrective and Preventive Actions (CAPA) have been completed.

This action has been finalised on 22/10/2026.
Action Commencement Date 7/04/2025
Responsible Entity Randox Australia Pty Ltd
Contact Information 02 9615 4640 - Randox Australia