TGA Action ID RC-2025-RN-00368-1
Type of Product Medical Device
Product Name/Description Human Care HC AB - Altair Lifting Solutions

ALTAIR 150 W/O HANGER BAR
Part number: 55200H

ALTAIR 220 W/O HANGER BAR
Part number: 55210H

ALTAIR 300 W/O HANGER BAR
Part number: 55220H
ARTG Number 168719
Action Type Product Alert
Action Level Consumer
Hazard Classification Class II
Reason for the Action Human Care HC AB have been informed of a fall incident caused by the failure to replace the belt annually as recommended in the inspection protocol and user manual. The User Manual specifies that the lift must be inspected at least once a year by a Human Care qualified personnel, and the lift belt should be replaced at the annual inspection.

Failure to follow the inspection instructions may lead to a broken/damaged lift belt and a potential for a patient fall.
Instructions for Customers/Users Customers are reminded to follow the User Manual and Annual Inspection Service Protocol, and ensure that the lift belts are replaced at the annual inspection.

Damaged/broken lift belts discovered during the daily inspection need to be replaced immediately as per the User Manual.

This action has been finalised on 8/7/2025.