| TGA Action ID |
RC-2025-RN-00359-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Turbo-Elite Laser Atherectomy Catheter
Model: 414-151, 417-152, 420-006, 423-001, 423-135-01, 423-135-02, 425-011, 425-135-01, 425-135-02
Size: 1.4 mm OTW, 1.7 mm OTW, 2.0 mm OTW, 2.3 mm OTW, 2.3 mm OTW, 2.3 mm OTW, 2.5 mm OTW, 2.5 mm OTW, 2.5 mm OTW |
| ARTG Number |
155552 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Philips has received complaints (approximately 0.01% occurrence rate) regarding the marker band (1.4-2.5 mm in size) detaching from the distal tip of the catheter and remaining in the patient as the laser catheter was retracted. This problem can occur if the epoxy did not form the locking feature between the marker band and fibre during manufacturing.
No patient harm has been reported in connection with any of the complaints received. In each scenario, the physician was able to detect the problem under fluoroscopy and successfully retrieve the marker bands using a balloon catheter.
UPDATE 16/07/2026: Philips has implemented corrective actions to address the manufacturing cause. Additionally, the IFU is being updated to reinforce fluoroscopic monitoring and to provide guidance on retrieval should detachment occur. |
| Instructions for Customers/Users |
Customers should be aware: - Philips recommends continuing to use the Turbo-Elite Laser Atherectomy Catheter as instructed, including using fluoroscopy to ensure that the marker band has not been left behind after the device has been retracted. It is also recommended to leave the guidewire in place until it has been confirmed that the marker band is still attached to the catheter after removal. - As a precaution, if a marker band detaches from the wire and remains in the patient, additional intervention is required to retrieve the marker band. There are multiple possible endovascular interventions to retrieve a detached marker band including snaring, balloon catheter inflation, and using a filter wire to capture the marker band. - If the above techniques are unsuccessful, then stenting can be used to secure the marker band in place to avoid distal embolization if it is in a big vessel. If it is in a small artery, it may be more appropriate to leave it there, as securing it with a stent could do more damage to the vessel. - Customers are requested to report all incidents of this problem to TGA and to Philips via a complaint.
UPDATE 16/07/2026: Customers are to: - Retain a copy of the Customer Letter with the product IFU until the revised IFU is available, and apply the updated warning to the use of the product. |
| Action Commencement Date |
7/04/2025 |
| Responsible Entity |
|
| Contact Information |
03 9525 5888 - Shannon Tolley, Life Systems Medical |