TGA Action ID RC-2025-RN-00356-1
Type of Product Medical Device
Product Name/Description CompoStop Platelet Storage Systems

Article Code: FD51600

Multiple Batch Numbers
ARTG Number 149449
Action Type Product Alert
Action Level Hospital
Hazard Classification Class II
Reason for the Action Fresenius Kabi has received 26 complaints with 43 confirmed defective units about leakages at the tube port. A leak in the sample bag could potentially lead to a microbial contamination in the product storage bag if undetected.

Investigation found this type of defect to be recognisable by the user under routine handling conditions, and that the likelihood a leakage leading to microbial contamination is improbable.
Instructions for Customers/Users Customers are to follow the IFU:
- When working with the CompoStop products
- In case of visible damage do not use the product
- Use a sample kit (P4206 - CompoFlex 1F 100ml sample system) to collect a new sample in case of a leak of the sample bag

No alternative stock is available currently. Fresenius will provide additional advice when alternative stock is expected to be available.
Action Commencement Date 1/04/2025
Responsible Entity Fresenius Kabi Australia Pty Ltd
Contact Information 1300 732 001 - Customer Service