| TGA Action ID |
RC-2025-RN-00352-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Hugo Robotic-Assisted Surgery (RAS) Sterile Interface Module (SIM)
Model number: MRASA0003 GTIN number: 10884521826564
Multiple serial numbers |
| ARTG Number |
380606 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
There's an increase in complaints related to intermittent connectivity of instruments with the Hugo RAS System.
This failure mode is defined as “Instrument ID failure” and occurs when the Hugo RAS System does not recognize an instrument being connected to the system. It can happen during setup of the Hugo RAS system and when in use intraoperatively. Investigation determined that connectivity issues can result intermittently from contact issue with the gold-plated pins of the Sterile Interface Module.
There have been 359 reported instances of this failure mode, of those, 77 procedures resulted in an injury.
Risks include extended procedure duration and/or discontinue use of the Hugo RAS system for the remainder of the procedure. Risk of bleeding and tissue damage/ trauma are also observed.
Update 30/06/2025: Medtronic has identified additional affected serial numbers of the Hugo RAS Sterile Interface Module (SIM) used with the Hugo RAS system. These additional serial numbers are of an older SIM design, which has the potential for intermittent connectivity. |
| Instructions for Customers/Users |
Customers are to: · immediately quarantine and discontinue use of the affected device · return the affected product for replacement stock · replace the SIM if this problem arises and report any incidents to the local Medtronic representative
Update 30/06/2025: Customers are to check the updated list of affected serial numbers and follow instructions in the customer letter to reconcile the recall process.
This action has been closed out on 26/05/2026 |
| Action Commencement Date |
31/03/2025 |
| Responsible Entity |
|
| Contact Information |
0402 897 512 - Maree Steffen, Medtronic Australasia |