TGA Action ID RC-2025-RN-00348-1
Type of Product Medical Device
Product Name/Description Welch Allyn Connex Devices

Welch Allyn Connex Vital Signs Monitor (CVSM)
Serial number: All
Multiple product codes

Welch Allyn CP150 Electrocardiograph
Serial number: All
Multiple product codes

Welch Allyn Connex Spot Monitor Accessory Power Management Stand (APM)
Serial number: All
Product code: 7000-APM

ARTG 220693, 327826, 340035
ARTG Number 220693, 327826, 340035
Unique Device Identifier (UDI) Refer to Attachment B and customer letter.
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Baxter Healthcare Corporation received reports of devices experiencing battery-related fires. Investigation found that the devices in question were utilising third-party, after-market batteries that are not approved by Baxter.

The use of an unapproved batteries may lead to combustion, fire or device performance problems.
Instructions for Customers/Users Baxter will update the IFUs and service manuals to include instructions to only use a Baxter approved battery for replacement and highlight the risks associated with using an unapproved battery.

Customers are to:
1. Use only approved batteries
2. Inspect all devices for the correct battery type
3. See attachment A in the customer letter for the updated IFU wording.

This action has been finalised on 9/11/2026.
Action Commencement Date 31/03/2025
Responsible Entity Welch Allyn Australia Pty Limited
Contact Information 1800 445 576 - Baxter Healthcare Customer Service