TGA Action ID RC-2025-RN-00341-1
Type of Product Medical Device
Product Name/Description CinchLock Flex Knotless Anchor

Catalogue number: CAT02643

GTIN number: 07613327123661

Multiple lot numbers
ARTG Number 229717
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action An increase in Pullwire breakages may result in the broken Pullwire remaining in the anchor body after deployment.
The potential risks of a Pullwire breakage include medical or surgical intervention or removal of a retained foreign body; retained foreign body resulting in inflammation or pain.

Risk Mitigation factors include:
• In some cases, it is still possible to remove the broken piece of the pullwire from the anchor completely.
• In cases where the broken pullwire cannot be removed from the anchor, it can be cut flush.
Instructions for Customers/Users Customers are to:

- Immediately dispose of all affected products.
- Contact their local Stryker representative to coordinate a replacement/credit of any affected product.
- Report any serious injuries, where the product has contributed or may have contributed to any adverse events, both to Stryker (Postmarketssp@stryker.com) and to the TGA.
Action Commencement Date 27/03/2025
Responsible Entity Stryker Australia Pty Ltd
Contact Information 0437 112 648 - Postmarketssp@stryker.com - Stryker