| TGA Action ID |
RC-2025-RN-00341-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
CinchLock Flex Knotless Anchor
Catalogue number: CAT02643
GTIN number: 07613327123661
Multiple lot numbers |
| ARTG Number |
229717 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
An increase in Pullwire breakages may result in the broken Pullwire remaining in the anchor body after deployment. The potential risks of a Pullwire breakage include medical or surgical intervention or removal of a retained foreign body; retained foreign body resulting in inflammation or pain.
Risk Mitigation factors include: • In some cases, it is still possible to remove the broken piece of the pullwire from the anchor completely. • In cases where the broken pullwire cannot be removed from the anchor, it can be cut flush. |
| Instructions for Customers/Users |
Customers are to:
- Immediately dispose of all affected products. - Contact their local Stryker representative to coordinate a replacement/credit of any affected product. - Report any serious injuries, where the product has contributed or may have contributed to any adverse events, both to Stryker (Postmarketssp@stryker.com) and to the TGA. |
| Action Commencement Date |
27/03/2025 |
| Responsible Entity |
|
| Contact Information |
0437 112 648 - Postmarketssp@stryker.com - Stryker |