TGA Action ID RC-2025-RN-00337-1
Type of Product Medical Device
Product Name/Description Alinity m System and Alinity m System Refurbished. An in vitro diagnostic medical device (IVD)

List numbers: 08N53-02, 08N53-32
ARTG Number 206870
Unique Device Identifier (UDI) (01)00884999048034, (01)00884999047587
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Abbott has identified that the System Solutions drawer enclosure design of the Alinity m System may allow liquid (liquid waste or system solution(s)) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.

If the leaked fluid (liquid waste or system solution(s)) accumulates within the System Solution drawer and potentially into the walking path of the user, the fluid could result in a physical hazard due to the potential for slip or fall. Additionally, leakage may introduce a potential for exposure to chemicals and potential biohazard exposure due to the presence of liquid waste originating from processed patient sample(s).

This problem has no impact on performance and clinical results generated. There have been zero occurrences of harm reported to date.
Instructions for Customers/Users An Alinity m System Liner will be installed to mitigate these leaks beyond the footprint of the instrument.

An Abbott representative will contact customers to schedule this installation. It is estimated the liner may be available around May 2025.

Customers are reminded to follow the Operations Manual, where users should always utilise Personal Protective Equipment when engaging with the instrument. In the event a leak is observed, take appropriate precautions to prevent exposure and immediately contact an Abbott Molecular Representative for additional troubleshooting information.

This action has been closed out on 02/12/2025.
Action Commencement Date 27/03/2025
Responsible Entity Abbott Australasia Pty Ltd Molecular Division
Contact Information 1800 816 696 - Abbott Customer Service