TGA Action ID RC-2025-RN-00335-1
Type of Product Medical Device
Product Name/Description CELL-DYN Ruby. An in vitro diagnostic medical device (IVD

Serial number: 70418BG
ARTG Number 199356
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action When expired reagents are scanned or manually entered on the CELL-DYN Ruby haematology analyser, the system changes the expiration date to current or future date. This change occurs without notifying the user that an expired reagent is inadvertently being used.

There have been no reports of impact to patient management.
Instructions for Customers/Users The CELL-DYN Ruby Operator’s Manual will be updated to provide information regarding this problem.

Customers are to:
- always use reagents within their valid expiration dates for all testing procedures.
- confirm the container labelling matches the data on the reagent log entry screen dialog box.
- familiarise with the updated Operator’s Manual regarding this problem.
Action Commencement Date 24/03/2025
Responsible Entity Abbott Australasia Pty Ltd Diagnostic Division
Contact Information 1800 816 696 - Abbott Customer Service