| TGA Action ID |
RC-2025-RN-00322-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Cartiva Synthetic Cartilage Implant devices
Cartiva Synethetic Cartilage Implant (SCI) 6MM Catalogue number: CAR06AUS Lot numbers: All
Cartiva Synthetic Cartilage Implant (SCI) 8MM Catalogue number: CAR08AUS Lot numbers: All
Cartiva Synthetic Cartilage Implant (SCI) 10MM Catalogue number: CAR10AUS Lot numbers: All |
| ARTG Number |
278815 |
| Action Type |
Recall / Product Alert |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Stryker has become aware of recently published data and post market reports indicating that patients implanted with Cartiva SCI may experience a higher-than-expected occurrence rate when compared to data submitted in the 2016 FDA PMA of the following documented hazards: revision, removal, implant subsidence, displacement, persistent or worsening pain, nerve damage or fragmentation.
Cartiva SCI devices have been observed in some cases to be revised/removed at higher rates than previously observed in the initial Cartiva SCI premarket and post-approval studies.
The labelling (IFU and surgical technique) includes warnings, precautions, and potential complications and adverse reactions related to the Cartiva product. The hazards listed are known potential hazards associated with these devices and related procedures. |
| Instructions for Customers/Users |
Hospitals are to: - Review the Customer Letter, quarantine then destroy any affected product - Inform Stryker if affected items has been transferred - Keep Stryker informed of any adverse events associated with this product
Surgeons are to: - Continue to follow patients treated with an impacted product for new or worsening symptoms of pain, difficulty walking, skin reactions, stiffness, swelling, or weakness of the joint, consistent with follow up protocols. - Per standard practice, continue to discuss all potential risks identified for Cartiva SCI and discuss the benefits and risks of all relevant treatment options with patients.
This action has been finalised on 9/12/2025. |
| Action Commencement Date |
12/08/2025 |
| Responsible Entity |
|
| Contact Information |
postmarketssp@stryker.com - Stryker |