TGA Action ID RC-2025-RN-00320-1
Type of Product Medical Device
Product Name/Description Philips DXR Systems – DuraDiagnost, DuraDiagnost Rel. 4

Serial Numbers: SN140052, SN150012, SN150017, SN150022, SN190050, SN190076

ARTG 207424 (cancelled 21/12/2023)
ARTG Number
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action Philips has identified a problem that may occur with the Philips DXR systems if the collimator is not attached correctly after repairing/replacing. The collimator may fall off during operation in the event of incorrect installation, which could result in injuries to patients and users.
Instructions for Customers/Users · Customers should be aware that during scan preparation, the tube assembly (Tube and Collimator) should be rotated to align with the Wallstand, set the SID at 110cm, and center the light field on the Wallstand detector (refer to figure 3.1 in the customer letter).
· Customers are required to note that if the laser from the collimator or crosshairs of the collimator's light field cannot align with the silkscreen on the front cover of the detector (refer to figure 3.2 in the customer letter), they should stop using the collimator and contact their local Philips representative to have it inspected.
· Customers should be aware that Philips will contact you to schedule a time for a Field Service Engineer to inspect the collimator to verify proper installation and make any necessary corrections.

This action has been closed out on 08/12/2025
Action Commencement Date 16/04/2025
Responsible Entity Philips Electronics Australia Ltd