| TGA Action ID |
RC-2025-RN-00302-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Ascenda Intrathecal Catheters
Model numbers: 8780, 8781, 8784
Manufactured: on or before 09 May 2024
ARTG's 192057, 192058, 192044 |
| ARTG Number |
192044, 192057, 192058 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
In May 2024, Medtronic notified their customers of a design update to the Model 8780, 8781, and 8784 AscendaTM Intrathecal catheters (Ascenda catheter).
The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.
As of 10-May-2024, all Ascenda catheters are manufactured with the updated design. At this time, Medtronic has sufficient inventory of the Ascenda catheters with the updated design and is recalling the prior configuration.
Medtronic is not recommending replacement of Ascenda catheter due to low occurrence rate of occlusion and the risks associated with replacement surgery. They advise patients and caregivers to be aware of signs and symptoms of withdrawal or underlying conditions to evaluate for potential catheter problems. |
| Instructions for Customers/Users |
Customers are to: ·Review all inventory of Ascenda Catheter(s) ·Identify the Ascenda catheter product manufactured on or before 09-MAY-2024 that does not have the updated design. ·Return affected product per the return instructions in the enclosed Customer Acknowledgement Form.?
This action has been closed out on 24/11/2025 |