TGA Action ID RC-2025-RN-00302-1
Type of Product Medical Device
Product Name/Description Ascenda Intrathecal Catheters

Model numbers: 8780, 8781, 8784

Manufactured: on or before 09 May 2024

ARTG's 192057, 192058, 192044
ARTG Number 192044, 192057, 192058
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action In May 2024, Medtronic notified their customers of a design update to the Model 8780, 8781, and 8784 AscendaTM Intrathecal catheters (Ascenda catheter).

The intent of the design update is to reduce the potential for tissue growth into the Ascenda catheter connector which may potentially lead to catheter occlusion.

As of 10-May-2024, all Ascenda catheters are manufactured with the updated design. At this time, Medtronic has sufficient inventory of the Ascenda catheters with the updated design and is recalling the prior configuration.

Medtronic is not recommending replacement of Ascenda catheter due to low occurrence rate of occlusion and the risks associated with replacement surgery. They advise patients and caregivers to be aware of signs and symptoms of withdrawal or underlying conditions to evaluate for potential catheter problems.
Instructions for Customers/Users Customers are to:
·Review all inventory of Ascenda Catheter(s)
·Identify the Ascenda catheter product manufactured on or before 09-MAY-2024 that does not have the updated design.
·Return affected product per the return instructions in the enclosed Customer Acknowledgement Form.?

This action has been closed out on 24/11/2025
Action Commencement Date 17/03/2025
Responsible Entity Medtronic Australasia Pty Ltd