TGA Action ID RC-2025-RN-00218-1
Type of Product Medical Device
Product Name/Description DxC 500 AU Clinical Chemistry Analyzer. An in vitro diagnostic medical device (IVD)

Reference numbers: C63519, C63520

Software versions: V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3

Multiple serial numbers
ARTG Number 177999
Unique Device Identifier (UDI) 14987666545058, 14987666545065
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action The analyzer may not run a requested calibration order for a two-part reagent and subsequently prevents any new calibration order from being placed.

For two-part reagents, when reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. The analyzer will not allow additional calibration orders to be requested for any assays.

If incorrect record in database is not corrected, it may cause delays in generating patient results.
Instructions for Customers/Users Customers are to:
- Not place calibration orders when the analyzer is processing samples.
- Refer to the DxC 500 AU IFU Reagent Overview section for instructions on Monitoring the Status of Reagents.
- Contact Beckman Coulter if you experience this problem on software version 1.3, 1.4, 1.4.1 or 1.4.2,

This action has been closed out on 06/08/2026
Action Commencement Date 11/03/2025
Responsible Entity Beckman Coulter Australia Pty Ltd
Contact Information 1800 060 881 - Beckman Coulter Customer Support Centre