| TGA Action ID |
RC-2025-RN-00218-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
DxC 500 AU Clinical Chemistry Analyzer. An in vitro diagnostic medical device (IVD)
Reference numbers: C63519, C63520
Software versions: V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3
Multiple serial numbers |
| ARTG Number |
177999 |
| Unique Device Identifier (UDI) |
14987666545058, 14987666545065 |
| Action Type |
Product Correction |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
The analyzer may not run a requested calibration order for a two-part reagent and subsequently prevents any new calibration order from being placed.
For two-part reagents, when reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. The analyzer will not allow additional calibration orders to be requested for any assays.
If incorrect record in database is not corrected, it may cause delays in generating patient results. |
| Instructions for Customers/Users |
Customers are to: - Not place calibration orders when the analyzer is processing samples. - Refer to the DxC 500 AU IFU Reagent Overview section for instructions on Monitoring the Status of Reagents. - Contact Beckman Coulter if you experience this problem on software version 1.3, 1.4, 1.4.1 or 1.4.2,
This action has been closed out on 06/08/2026 |
| Action Commencement Date |
11/03/2025 |
| Responsible Entity |
|
| Contact Information |
1800 060 881 - Beckman Coulter Customer Support Centre |