| TGA Action ID |
RC-2025-RN-00182-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Comprehensive Shoulder Stems 5mm Micro, 6 mm Micro 7 mm Micro and 7 mm Mini
Material numbers: 113605, 113606, 113607, 113627
Batch numbers: 66173215, 66173219, 66173220, 65915809
ARTG 262959 & 227906 |
| ARTG Number |
227906, 262959 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Five global complaints reported that the stem implant did not fit in the inserter or that the inserter could not be detached from the stem.
There were no injuries reported associated with the five complaints received. The internal investigation identified excess metal on the inner taper of the stem due to a chipped manufacturing tool.
If the problem occurs the surgeon may need to find another device which can lead to an extension of the surgery.
There is no impact to patients with the previously implanted devices. |
| Instructions for Customers/Users |
Customers are to: - Inspect and quarantine affected stocked. - Contact a Zimmer Biomet Sales Representative to return affected stock. - Contact customer service to arrange for replacement stock.
This action has been closed out on 22/07/2025. |
| Action Commencement Date |
7/03/2025 |
| Responsible Entity |
|
| Contact Information |
0481 905 548 - Jasmine Khor, Zimmer Biomet |