Type of Product Medical Device
Product Name/Description Comprehensive Shoulder Stems 5mm Micro, 6 mm Micro 7 mm Micro and 7 mm Mini

Material numbers: 113605, 113606, 113607, 113627

Batch numbers: 66173215, 66173219, 66173220, 65915809

ARTG 262959 & 227906
ARTG Number 227906, 262959
Action Type Recall
Action Level Hospital
Hazard Classification Class II
Reason for the Action Five global complaints reported that the stem implant did not fit in the inserter or that the inserter could not be detached from the stem.

There were no injuries reported associated with the five complaints received. The internal investigation identified excess metal on the inner taper of the stem due to a chipped manufacturing tool.

If the problem occurs the surgeon may need to find another device which can lead to an extension of the surgery.

There is no impact to patients with the previously implanted devices.
Instructions for Customers/Users Customers are to:
- Inspect and quarantine affected stocked.
- Contact a Zimmer Biomet Sales Representative to return affected stock.
- Contact customer service to arrange for replacement stock.

This action has been closed out on 22/07/2025.