| TGA Action ID |
RC-2025-RN-00161-1 |
| Type of Product |
Medical Device |
| Product Name/Description |
Delivery cable Flex-Pusher II (51FP100) in combination with Delivery system ODSIII 6F (98DS006 & 98US006) and ODSv1 6F (51US006)
Flex-Pusher II (51FP100) is part of the Procedure Pack Reference number: 29ASD06F, 29ASD07F, 29ASD09F, 29ASD10F, 71VSD06F, 89VSD08SF, 84VSD08F, 90VSD08LF
ARTG's: 203602 & 421456 |
| ARTG Number |
203602, 421456 |
| Action Type |
Recall |
| Action Level |
Hospital |
| Hazard Classification |
Class II |
| Reason for the Action |
Certain Procedure Packs containing Flex-Pusher II (51FP100) have a size recommendation of 6F for ODSIII and ODSv1 on the label of the Procedure Pack box and the sterile pouch of the implant that might lead to incompatibility.
Compatibility problems between the Flex-Pusher II delivery cable and the ODS (III or v1) Delivery Set may arise due to variations in their respective diameters within specified tolerance limits.
To date, there have been four complaints associated with this incorrect compatibility. No patient adverse events or injury have been reported. |
| Instructions for Customers/Users |
Customers should check their inventory and quarantine any of the affected procedure packs. These should then be returned to the local Occlutech representative.
Affected products will be replaced by corrected items.
This action has been finalised on 8/7/2025. |
| Action Commencement Date |
13/03/2025 |
| Responsible Entity |
|
| Contact Information |
info.mdso@occlutech.com - Occlutech |