TGA Action ID RC-2025-RN-00156-1
Type of Product Medical Device
Product Name/Description Centricity Universal Viewer (UV), Centricity Radiology RA600, & Centricity Cardiology CA1000

Multiple product names, software versions and GTIN numbers.

ARTG's: 161079 & 246996
ARTG Number 161079, 246996
Action Type Product Correction
Action Level Hospital
Hazard Classification Class II
Reason for the Action GE HealthCare has become aware of a potential security vulnerability in Centricity Universal Viewer (UV), Centricity Radiology RA600 (RA600), & Centricity Cardiology CA1000 (CA1000) where the service login credentials are able to be identified, which could allow a malicious actor with these credentials to access the system and potentially manipulate patient data.

There have been no injuries or unauthorised access to patient data reported to GE HealthCare because of this problem.
Instructions for Customers/Users Customers can continue to use their device, ensuing all potential users are made aware of this safety notification.
Customers should:
1: Ensure their hospital network is secure
and
2: Outside of the hospital network implement additional safeguards such as a VPN.
GE HealthCare will provide a software correction and will contact customers to arrange for the correction.

Update 16/05/2025:GE HealthCare has released an installation manual addendum that provides guidance for customers on updating their system passwords.
The update includes:
1. Instructions for Accessing the Addendum: Step-by-step directions to download the installation manual addendum.
2. Password Change Procedures: Detailed instructions within the addendum outlining how customers can securely change their system passwords.

This action has been closed out on 09/06/2026
Action Commencement Date 11/03/2025
Responsible Entity GE Healthcare Australia Pty Ltd
Contact Information 1800 659 465 - GE HealthCare Service