| Reason for the Action |
Two failure modes may be experienced during use:
1. A bent or detached heater wire 2. Silicone peeling or detaching from the Jaws of the Harvesting Tool.
Peeling or detachment of silicone from the Jaws can cause the tool to malfunction. If the Jaw Heater Wire becomes compromised, bent, or detached, it could potentially harm the patient:
-Procedure delays, retrieve detached silicone, or set up a new device. -Additional interventions to retrieve detached material, complete the vessel harvest, address bleeding, and/or harvest additional conduit. -Bleeding due to incomplete cauterization or damage. -Burns to vessel, conduit, or tissue from damaged Jaw material. -Retained detached silicone within the body cavity.
If unnoticed or not removed, detached silicone can cause future complications, including delayed pain, infection, localised allergic reactions, and/or readmission for further medical or surgical intervention. |